Motion characterization scheme to minimize motion artifacts in intravital microscopy

  • Lee, Sungon; 
  • Courties, Gabriel; 
  • Nahrendorf, Matthias; 
  • Weissleder, Ralph; 
  • Vinegoni, Claudio
Citations

WEB OF SCIENCE

17
Citations

SCOPUS

18

초록

Respiratory- and cardiac-induced motion artifacts pose a major challenge for in vivo optical imaging, limiting the temporal and spatial imaging resolution in fluorescence laser scanning microscopy. Here, we present an imaging platform developed for in vivo characterization of physiologically induced axial motion. The motion characterization system can be straightforwardly implemented on any conventional laser scanning microscope and can be used to evaluate the effectiveness of different motion stabilization schemes. This method is particularly useful to improve the design of novel tissue stabilizers and to facilitate stabilizer positioning in real time, therefore facilitating optimal tissue immobilization and minimizing motion induced artifacts. (C) 2017 Society of Photo-Optical Instrumentation Engineers (SPIE).

키워드

motion artifact; intravital microscopy; motion compensation; in vivo imaging; IN-VIVO; RESOLUTION; HEART
제목
Motion characterization scheme to minimize motion artifacts in intravital microscopy
저자
Lee, Sungon; Courties, Gabriel; Nahrendorf, Matthias; Weissleder, Ralph; Vinegoni, Claudio
DOI
10.1117/1.JBO.22.3.036005
발행일
2017-03
유형
Article
저널명
Journal of Biomedical Optics
권
22
호
3
페이지
1 ~ 8