자발 구동형 종이 펌프를 이용한 종이 전기화학 센서

Paper-based Electrochemical Sensor Using a Self-operated Paper Pump
  • Hua, S.H.
  • Kim, C.
  • Nguyen, D.C.
  • Kim, Y.S.
Citations

SCOPUS

0

초록

We developed a self-operated paper pump that can maintain a nearly constant flow rate of an aqueous solution along a paper strip channel in paper-based analytical devices (PADs). The quasi-stationary flow rate was controlled by increasing the crosssectional channel area (capillary force) using a fan-shaped absorption pad coupled with a paper strip channel. The flow rate is regulated by varying the fan angle of the circular absorbing pad. Furthermore, the flow rate can be increased by furnishing a hollow cavity at the center of a conventional paper strip channel. The rate was regulated by varying the length of the hollow paper channel in the flow rate range of 5.1-26.4 mm/min. As a preliminary work, a paper-pump-coupled PAD was fabricated, and its CV detection capability was evaluated for the redox reaction of Fe(CN)6 +4/+3. The combination of a paper pump with a PAD resulted in an ideal CV curve with a higher limiting current and faster response time. These results are interpreted well by the Levich equation, which suggests that the paper pump is a very useful component in paper-based sensors. © 2024, J. Sens. Sci. Technol. All rights reserved.

키워드

CV measurementPADPaper pumpPaper-based electrochemical sensor
제목
자발 구동형 종이 펌프를 이용한 종이 전기화학 센서
제목 (타언어)
Paper-based Electrochemical Sensor Using a Self-operated Paper Pump
저자
Hua, S.H.Kim, C.Nguyen, D.C.Kim, Y.S.
DOI
10.46670/JSST.2024.33.3.158
발행일
2024-05
유형
Article
저널명
센서학회지
33
3
페이지
158 ~ 164