상세 보기
Evaluating QoS of a wireless system for real-time cardiac monitoring
- Park, Juyoung;
- Lee, Jaemyoun;
- Ryu, Junhee;
- Shin, Heonshik;
- Heu, Shin;
- ... Kang, Kyungtae
WEB OF SCIENCE
3SCOPUS
3초록
Quality of service (QoS), and in particular reliability and low latency, are essential qualities of safety-critical wireless systems for medical applications. However, wireless links are typically prone to bursts of errors, with characteristics which vary over time. We propose a wireless system suitable for real-time remote patient monitoring in which both the necessary reliability and a bounded end-to-end delay are achieved by an efficient error control scheme. We have paired an example electrocardiography (ECG) application to this wireless system. We also developed a tool chain to assess various combinations of system parameters: we can determine the QoS in terms of the packet-delivery ratio and the end-to-end latency, and also the size of jitter buffer required for seamless ECG monitoring services. A realistic assessment, based on data from the MIT-BIT arrhythmia database, shows that the proposed wireless system can achieve an appropriate level of QoS for real-time ECG monitoring if link-level error control is appropriately implemented. © 2013 IEEE.
키워드
- 제목
- Evaluating QoS of a wireless system for real-time cardiac monitoring
- 저자
- Park, Juyoung; Lee, Jaemyoun; Ryu, Junhee; Shin, Heonshik; Heu, Shin; Kang, Kyungtae
- 발행일
- 2013-03
- 유형
- Conference Paper
- 저널명
- Proceedings - International Conference on Advanced Information Networking and Applications, AINA
- 페이지
- 1105 ~ 1112
- 언어
- ENG
- 분량
- 8 페이지
- ISSN
- E 1550-445X
P 1550-445X