Evaluation of peristaltic micromixers for highly integrated microfluidic systems

  • Kim, Duckjong
  • Rho, Hoon Suk
  • Jambovane, Sachin
  • Shin, Soojeong
  • Hong, Jong Wook
Citations

WEB OF SCIENCE

1
Citations

SCOPUS

4

초록

Microfluidic devices based on the multilayer soft lithography allow accurate manipulation of liquids, handling reagents at the sub-nanoliter level, and performing multiple reactions in parallel processors by adapting micromixers. Here, we have experimentally evaluated and compared several designs of micromixers and operating conditions to find design guidelines for the micromixers. We tested circular, triangular, and rectangular mixing loops and measured mixing performance according to the position and the width of the valves that drive nanoliters of fluids in the micrometer scale mixing loop. We found that the rectangular mixer is best for the applications of highly integrated microfluidic platforms in terms of the mixing performance and the space utilization. This study provides an improved understanding of the flow behaviors inside micromixers and design guidelines for micromixers that are critical to build higher order fluidic systems for the complicated parallel bio/ chemical processes on a chip. (C) 2016 AIP Publishing LLC.

키워드

CHIPMICROPUMPSVALVESPUMP
제목
Evaluation of peristaltic micromixers for highly integrated microfluidic systems
저자
Kim, DuckjongRho, Hoon SukJambovane, SachinShin, SoojeongHong, Jong Wook
DOI
10.1063/1.4940927
발행일
2016-03
유형
Article
저널명
Review of Scientific Instruments
87
3
페이지
1 ~ 9